Chinese Journal of Magnetic Resonance ›› 2004, Vol. 21 ›› Issue (3): 337-344.

Previous Articles     Next Articles

DESIGN OF A MAGNET USED IN A PORTABLE NMR ROCK ANALYZER

 SUN Wei1, WANG Wei-Min2, GU Chang-Chun1, ZHANG Yan-Lei3   

  1. 1.Institute of porous Flow and Fluid Mechanics,  CNPC and the Chinese Academy of Sciences, Langfang 065007, China;  2.Quantum information and measurement Key Laboratory the Ministry of Education Peking University, Beijing 100871, China; 3.No.1 Drilling Engineering Co. of Huabei Petroleum Administration Bureau CNPC, Langfang 065007, China
  • Received:2004-02-19 Revised:2004-04-29 Online:2004-09-05 Published:2004-09-05

Abstract:

There are many ways of design for the magnets used in nuclear magnetic resonance. However,  it is always challenging to make the magnet small, light, stable and having large volume with uniform magnetic field. Solution to this challenge lies in accurate computation of the electromagnetism field. In this paper, we first briefly reviewed the problems in magnet design, and then showed how to optimize the different ways of magnet design. An optimized magnet design was proposed, based on which a portable NMR cock analyzer was built successfully. The results from a number of on site tests showed that the portable NMR rock analyzer we built has high stability and good reproducibility, and is suitable for working under severe conditions.

Key words: NMR, permanent magnet, electromagnetic field, stability, reproducibility

CLC Number: