DEVELOPMENT OF L BAND THREE DIMENSIONAL ESR IMAGING SYSTEM(Ⅴ)——System Configuration and Performance Evaluation

WU Ke1*, CONG Jian-Bo1, ZHANG Qing-Jun1, XIAN Hong1, WANG Chang-Zhen1, SUN Cun-Pu1, ZHENG Ying-Guang2, XU Jing2, DONG Feng-Xia2, SHEN Er-Zhong2, XU Yu-Shu2   

  1. 1.Institute of Radiation Medicine, Academy of Military Medical Science, National Center of Biomedical Analgsis, FR&ESR Lab, Beijing 100850, China; 2.Analytic and Measure Center, Jilin University, Changchun 130022, China
  • Received:2003-07-24 Revised:2003-09-29 Online:2004-06-05 Published:2004-06-05
  • Supported by:

    国家自然科学基金资助项目(29935080)

Abstract:

The detection sensitivity and the achievable spatial resolution of a newly developed L-band three-dimensional ESR imaging system were evaluated. The detection sensitivity was measured using a group of samples of aqueous Tempo solution. A signal to noise ratio (S/N) of about 4∶1 was obtained on a cylindric al sample with diameter of 10 mm and height of 30 mm, which corresponds to the minimum detectable concentration of 1×10-4 mol/L. To obtain projections with sufficient S/N for imaging, a higher concentration of more than 5×10-4 mol/L and a larger sample size (diameter of φ19 mm and height of 30 mm) was needed. Measured on solid DPPH samples, the spatial resolution of imaging can reach about 1 mm. The general performance features of the ESRI system were also discussed in the paper.

Key words: ESR, 3D-ESRI, sensitivity, resolution, performance feature