The continuous wave and pulse resonant cavities were designed and manufactured for the pulsed electron paramagnetic resonance (EPR) spectrometer. The resonant cavity used for the continuous wave EPR was a rectangular resonator, and the pulse resonant cavity adopted the structure of planar microresonator. The software of 3D electromagnetic simulation, Ansoft-HFSS, were used for simulation. The parameters of the designed resonators were evaluated with the network analyzer. The experimental data were found to be consistent with those of theoretical simulation.