极化3He系统核磁共振FID信号的激光探测
Laser Detection of Nuclear Magnetic Resonance FID Signal for Polarized 3He System
通讯作者: 陈思宇, Tel: 0816-2487741, E-mail:chensy_nemail@163.com
收稿日期: 2022-03-21
基金资助: |
|
Received: 2022-03-21
极化率测量装置是极化3He系统的重要组成部分之一.本文介绍了一套在极化3He系统上搭建的激光探测自由感应衰减核磁共振(FID NMR)信号的实验装置,该装置有望为3He极化率的高精度实时测量提供新的途径.激光探测法的原理是法拉第旋光效应,该方法使用一束线偏振光探测极化3He磁矩绕主磁场进动的FID信号,实验结果表明相较于拾波线圈探测方法,激光探测方法的信噪比提高了106%,对3He的极化率测量更精确.该探测方法有望替代传统的拾波线圈,广泛地应用于3He极化系统的极化率测量或精密测量实验中.
关键词:
Polarimetry is an important part of a polarized 3He system. This work describes a laser detection setup which measures the free induction decay nuclear magnetic resonance (FID NMR) signal in the spin-polarized 3He system. The setup provides a novel approach to the high precision real-time polarization measurement of the polarized 3He systems. This laser detection method is based on Faraday rotation effect, which utilizes a linearly polarized laser beam to detect the FID signal of the magnetic moment of 3He atoms which precess along the direction of the main magnetic field. Compared with traditional pick-up coils, experimental results demonstrate that the laser detection method has a signal-to-noise ratio improvement of 106% and consequently higher precision in 3He polarization measurement. The laser detection method proposed in this paper can replace the pickup coils, and it has various applications in the polarization measurement of 3He systems or in precision measurement experiments.
Keywords:
本文引用格式
王昶沣, 兰阳, 闫海洋, 彭梅, 陈思宇.
WANG Chang-feng.
引言
自旋交换光泵浦(Spin Exchange Optical Pumping,SEOP)技术是常用的极化3He的方法,这也是本文采用的极化方法.该方法通过光泵浦极化碱金属Rb的价电子,碱金属原子再通过超精细相互作用将角动量传递给3He原子核,实现3He的极化.极化率是极化3He系统的核心指标之一,通常可以采用绝热快速(Adiabatic Fast Passage,AFP)NMR和电子顺磁共振(Electron Paramagnetic Resonance,EPR)相结合的技术测量3He的绝对极化率,也可以通过自由感应衰减(Free Induction Decay,FID)NMR信号测量其相对极化率,其中后者常用作3He原子极化状态的实时监测方法.3He极化磁矩和Rb极化磁矩的相互作用可以用偶极磁场表示,如(1)式:
其中
通过FID NMR技术探测3He相对极化率时,FID信号的幅值与极化率呈线性正相关.目前常见的探测方式为利用磁感应效应通过拾波线圈探测FID NMR信号,其探测对象为经典偶极磁场.在中国绵阳研究堆(China Mianyang Research Reactor,CMRR)[14, 15]的首个用于中子极化的极化3He系统采用拾波线圈的方法实现了极化率的实时监测[16],其3He极化率可以达到72.7%±0.4%[17].这种探测方式结构简单、成本较低、对极化率的损失较小.但由于其探测的经典偶极磁场以
其中
本文在法拉第旋光效应的基础上搭建了一套极化率激光探测装置,该装置被应用在一套基于SEOP方法的极化3He系统上,实验结果表明该方法实现了对3He原子相对极化率的测量.
1 实验部分
1.1 实验原理
本文使用的激光探测方法基于法拉第旋光效应:一束线偏振光可分解为左旋和右旋圆偏振光,当光传播方向上有电子极化时,极化电子对左右旋分量的折射率不同,从而引起光偏振面的旋转,即圆双折射效应.因此当探测光通过3He气室后,由左旋和右旋圆偏振光组成的线偏振光的偏振面会发生偏转,并且该偏转角度与极化率呈正比.
图1
图1
激光探测方法示意图. (a) 3He极化后的自旋方向朝向主磁场方向;(b)在垂直于主磁场方向施加一射频场使3He偏离主磁场一个较小的角度;(c) 3He自旋弛豫过程;(d)极化3He恢复到初始状态
Fig.1
Illustration of the laser detection method. (a) The polarized 3He has spin aligned to the direction of the main magnetic field; (b) An radiofrequency electromagnetic field is applied in a direction perpendicular to the main magnetic field to tip the spin of the 3He a small angle away from the main magnetic field; (c) The process of 3He spin relaxation; (d) The polarized 3He is back to its original status
上式中,α为宏观磁矩偏离主磁场方向的夹角,α的大小与施加射频的频率、振幅和脉冲时间有关;f为3He拉莫尔进动频率;φ为进动的初始相位;
沿着气室的x轴方向入射一束偏离Rb的D1线的线偏振光,根据法拉第旋光效应,其偏转角会随着Rb原子感受到的磁场的变化而变化.探测过程中宏观磁矩
其中,l为探测光在气室中的光程,
1.2 实验装置
为了实现3He的极化,作者及所在团队搭建了如图 2所示的极化3He装置[16],该装置也是国内首个用于中子极化的装置.装置可分为四个子系统:光泵系统、加热系统、磁场环境和3He极化率检测系统.装置的核心是一个圆柱形气室,由GE180玻璃制成,其直径为6.5 cm、长为6 cm.气室中封装了约1.16 bar(1 bar = 105 Pa)的3He气体(常温)、50 mbar的N2(缓冲气体)和少量(< 0.1 mg)自然丰度的Rb.由于常温下Rb原子为固态,因此实验时需要对气室进行加热使其变成Rb蒸汽.加热炉由三层聚二醚酮(Polyether-ether-ketone,PEEK)材料制成,其四周有四个双层透明的熔融石英玻璃,该玻璃对794.7 nm激光的透过率为99.9%.气室放置于加热炉中心,使用热的压缩空气将其缓慢加热到170 ℃.加热炉的升温过程由自制温控箱控制,温度精度为0.1 ℃.主磁场线圈为方形四线圈构型的Merrit线圈,该线圈可以沿z轴方向提供一个强度约为16 G(1 G = 10−4 T)的恒定磁场,对应的3He原子核的拉莫尔进动频率为52 kHz.本文采用半导体二极管阵列式激光器沿z轴方向提供波长为794.7 nm、线宽为0.15 nm、功率为100 W的圆偏振光泵浦Rb的D1线,然后被极化的Rb原子通过超精细相互作用将其电子的自旋传递给3He原子核,从而实现3He的极化.
图2
为验证本文所提的激光检测方法的有效性,选择了与拾波线圈测量方法进行对比.拾波线圈是一种可以实时监测3He原子相对极化率的方法:一对拾波线圈沿x轴置于气室的两侧,其平均半径为5.2 cm、电阻值为614.4 Ω.该方法由于易于实现而被广泛使用,但其信噪比受到线圈参数、与气室的距离等多因素的影响,优化工作比较困难.
本文使用的激光探测方法使用一束线偏振光沿x轴经过气室从而探测FID NMR信号,其探测光路如图 3所示.分布布拉格反射镜(Distributed Bragg Reflector,DBR)激光器(Probe laser)沿x轴方向提供了一束功率为50 mW的线偏振光,经非球面透镜(Lens,焦距f = 6 mm)准直后变成一束平行光.光路中的光隔离器(Optical Isolator)是为了防止其他杂散光反射从而影响激光的稳定性.光隔离器后分别是半波片和偏振分束镜(Polarizing Beam Splitter,PBS),它们组合使用可以对入射进气室的激光能量进行调节,同时在偏振分束镜之后加入了激光振幅稳定器(Laser Amplitude Stabilizer)可将激光器的功率稳定至±0.001 mW.激光经半波片调节偏振方向后进入气室.经过气室后的激光经PBS分解成偏振态相互垂直的两束线偏振光,用平衡探测器(Balanced Detector)的两个光敏面分别将两束光信号直接转化为电信号,最后使用数据采集卡(Data Acquisition,DAQ)采集电信号到计算机内部进行处理.
图3
为了对结果进行对比,本文同时采用了拾波线圈探测和激光探测FID NMR信号的方法,其中拾波线圈既作为发射线圈,也作为接收线圈,实验装置如图 4所示.
图4
1.3 实验过程
控制系统通过DAQ给拾波线圈发送振幅为2.5 V、频率f为52.23 kHz、脉冲宽度为2.5 ms的射频脉冲,方向为x轴方向,之后再通过DAQ同时采集拾波线圈和平衡探测器得到的FID原始数据.由于系统中存在各类电子噪声,为了得到高频小信号,本文采用相敏检测方法对FID原始数据进行处理.
FID信号可表示为:
其中信号振幅A与3He极化率成正比,
其中
然后对数据进行积分实现信号提取:
其中N是被积信号周期的个数,T表示被积信号的周期.从(7)和(8)式可以看出,原信号与参考信号相乘之后,系统噪声的频谱也被搬移至f处,(9)式和(10)式表示的积分相当于对数据进行了低通滤波处理,滤除掉信号中的高频成分,仅留下差频部分,以此达到提取小信号的目的.本文使用了一个4阶巴特沃斯滤波器滤除高频成分和噪声成分,最后得到差频信号,并采用(5)式对FID NMR信号进行拟合.
2 结果与讨论
最终得到的拾波线圈和激光探测方法的输出信号的x分量分别如图 5所示.拾波线圈探测的FID NMR信号的初始振幅为88.180±0.64 μV,频率为149.670±0.036 Hz,横向弛豫时间T2为49.29±0.51 ms;激光探测的FID NMR信号的初始振幅为1.370±0.007 mV,频率为151.923±0.024 Hz,横向弛豫时间T2为74.02±0.82 ms.
图5
图5
(a) 拾波线圈探测的FID NMR信号,蓝色表示拾波线圈采集到的信号(放大倍数为100倍),红色表示拟合曲线;(b)激光探测的FID NMR信号,蓝色表示平衡探测器采集到的信号,红色表示拟合曲线
Fig.5
(a) FID NMR signal acquired with pickup coil, the blue represents the signal collected by the pickup coil (the magnification is 100 times), and the red represents the fitting curve; (b) FID NMR signal acquired with laser detection method, the blue represents the signal collected by the balanced detector, and the red represents the fitting curve
从探测结果可以发现:(1)利用拾波线圈和激光同时探测FID NMR信号,激光探测得到的FID NMR信号具有更长的横向弛豫时间T2.这是因为T2主要与气室、压力、主磁场梯度等因素有关.使用拾波线圈探测的是整个气室宏观磁场的平均值,而激光探测的是探测光路上的平均磁场,探测光路上的磁场梯度小于整个气室的磁场梯度,所以激光探测的T2更长;(2)从拟合数据中还可以得到拾波线圈探测的频率为149.670±0.036 Hz,激光探测的频率为151.923±0.024 Hz.造成频差的原因可能有以下三点:①激光探测的磁场为费米接触式,对于球形气室,经典偶极磁场在气室中产生的磁场为0,但是在一般形状的气室中,经典偶极磁场在气室中的贡献不为0,由此导致激光探测的信号中存在经典偶极磁场的贡献;②拾波线圈探测的是整个3He气室宏观磁场的平均值,探测的信号与线圈的几何形状、共振性质(电阻,感抗)、以及与气室远近(相对几何位置)有关,而激光探测的是探测光路上平均的磁场;③主磁场的不均匀性.
在频域下对FID NMR信号进行信噪比分析(图 6),图中红色虚线部分用于计算噪声.可以得到拾波线圈探测的信噪比为24.75,激光探测的信噪比为50.98.激光探测的信噪比约是拾波线圈的2倍,这是因为拾波线圈探测的是整个气室宏观磁场的平均值,而探测光探测的是探测光路上Rb的磁场.探测光的直径为4 mm,气室直径为65 mm,相当于探测光只与气室中少量的Rb原子相互作用,因此当增加探测光光束直径,光探测的信噪比有望进一步提高.
图6
图6
(a) 拾波线圈探测的FID NMR信号频谱图;(b)激光探测的FID NMR信号频谱图.图中红色虚线的部分用于计算噪声
Fig.6
(a) Power spectrum density of FID NMR signal acquired with pickup coil; (b) Power spectrum density of FID NMR signal acquired with laser detection method. The red dotted lines in the figures are used for calculating the noise
3 结论
本文对激光探测FID NMR信号的原理进行了简要的介绍,并基于极化3He系统搭建了一套极化率激光探测装置.最后对拾波线圈和激光探测FID NMR信号进行了理论和实验上的比较,激光探测法的信噪比更高、3He极化率测量更精确.激光探测法探测的FID NMR信号也具有更长的横向弛豫时间T2,为之后的精密测量做下了铺垫.为了进一步优化激光探测装置,提高光探测的信噪比,可以采用提高加热炉的温度或对探测光进行扩束以增加与探测光相互作用的Rb原子数的方式.
无
参考文献
Comparison of in vivo lung morphometry models from 3D multiple b-value 3He and 129Xe diffusion-weighted MRI
[J]. ,DOI:10.1002/mrm.27608 [本文引用: 1]
Lung MRI with hyperpolarised gases: current & future clinical perspectives
[J]. ,
3He diffusion MRI in human lungs
[J]. ,
Three-dimensional isotropic functional imaging of cystic fibrosis using oxygen-enhanced MRI: Comparison with hyperpolarized 3He MRI
[J]. ,DOI:10.1148/radiol.2018181148 [本文引用: 1]
Probing the short range spin dependent interactions by polarized 3He atom beams
[J]. ,DOI:10.1140/epjc/s10052-014-3088-8 [本文引用: 1]
Searching for new spin- and velocity-dependent interactions by spin relaxation of polarized 3He gas
[J]. ,DOI:10.1103/PhysRevLett.115.182001
Search for spin-dependent short-range force using optically polarized 3He gas
[J]. ,DOI:10.1103/PhysRevD.85.031505 [本文引用: 1]
Polarization analysis on the LET cold neutron spectrometer using a 3He spin-filter: first results
[J]. ,DOI:10.1088/1742-6596/1316/1/012007 [本文引用: 1]
Development and application of a 3He neutron spin filter at J-PARC
[J]. ,DOI:10.1016/j.nima.2020.164301
Polarization analysis for small-angle neutron scattering with a 3He spin filter at a pulsed neutron source
[J]. ,
In-situ 3He SEOP polarizer for thermal neutrons
[J]. ,DOI:10.1088/1742-6596/1316/1/012009
3He neutron spin filter cell development program at JCNS
[J]. ,DOI:10.1088/1742-6596/528/1/012015 [本文引用: 1]
3He neutron spin filter at ILL
[J]. ,
A new operating neutron scattering facility CMRR in China
[J]. ,DOI:10.1080/10448632.2016.1233018 [本文引用: 1]
光泵抽运3He极化程度监控系统的设计与实现
[J]. ,
Design and implementation of a monitoring system for measuring polarization degree of optical pumping 3He
[J].
Development of a polarized 3He neutron spin filter based on spin exchange optical pumping at China Mianyang Research Reactor
[J]. ,DOI:10.1007/s11433-019-9410-3 [本文引用: 1]
Dynamics of two overlapping spin ensembles interacting by spin exchange
[J]. ,DOI:10.1103/PhysRevLett.89.253002 [本文引用: 1]
Collisions give sense of direction
[J]. ,DOI:10.1038/nphys204 [本文引用: 1]
Accurate 3He polarimetry using the Rb Zeeman frequency shift due to the Rb−3He spin-exchange collisions
[J]. ,DOI:10.1103/PhysRevA.58.3004 [本文引用: 1]
A frequency determination method for digitized NMR signals
[J]. ,
Feasibility study of a 3He-magnetometer for neutron electric dipole moment experiments
[J]. ,DOI:10.1016/S0168-9002(99)01024-4 [本文引用: 1]
3He maser for earth magnetic field measurement
[J]. ,DOI:10.1063/1.1605494 [本文引用: 1]
Search for axion-like dark matter with spin-based amplifiers
[J]. ,DOI:10.1038/s41567-021-01392-z [本文引用: 1]
/
〈 | 〉 |