Chinese Journal of Magnetic Resonance ›› 2012, Vol. 29 ›› Issue (4): 508-520.

Previous Articles     Next Articles

Gradient Coil of Permanent Magnet Mini-Magnetic Resonance Imager and Image Quality

 HOU Shu-Lian1*, XIE Huan-Tong2, HOU Xiao-Wen3, LI Shi-Yu1, CHEN Wei1   

  1. 1. College of Elementary Medicine, Hebei United University, Tangshan 063000, China;
    2. Shanghai HuanTong Science and Education Equipment Co., LTD, Shanghai 201806, China;
    3. Hangzhou Dianzi University, Hangzhou 310000, China
  • Received:2011-11-18 Revised:2012-05-08 Online:2012-12-05 Published:2012-12-05
  • Supported by:

    河北省科学技术厅科研计划资助项目(08202133D).

Abstract:

Effects of gradient field linearity on the quality of images acquired on a permanent magnet mini-magnetic resonance imager were analyzed by theoretical analyses and numerical simulation. It was shown that gradient field with high linearity is a prerequisite to obtain high-quality images of small animals, such as mice, on permanent magnet, low field strength micro-magnetic resonance imagers. The gradient field linearity specification reported in the literature is too low to acquired high-quality mouse images. The image quality could not be improved by software compensation methods. The newly developed gradient field coil with high linear property and our independent intellectual property was also presented. The optimized gradient coil has been installed on our instrument, and yielded a better quality image.

Key words: MRI, Permanent Magnet Mini-mgnetic 3D Resonance Imager, linearity, gradient coil, simulation