Chinese Journal of Magnetic Resonance ›› 2018, Vol. 35 ›› Issue (2): 141-149.doi: 10.11938/cjmr20172596

Previous Articles     Next Articles

Optimization of Selective Radio Frequency Pulses for Simultaneous Multi-Slice MRI

LU Shan1,2, CHANG Yan2, QIAN Song-song1, SHI Bo2, YANG Xiao-dong2   

  1. 1. College of Electrical Engineering and Optoelectronic Technology, Nanjing University of Science and Technology, Nanjing 210094, China;
    2. Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou 215163, China
  • Received:2017-09-30 Online:2018-06-05 Published:2017-12-27

Abstract: Simultaneous multi-slice magnetic resonance imaging (MRI) has attracted much research interests recently due to its advantages in increasing signal-to-noise ratio (SNR) and decreasing acquisition time. To achieve simultaneous multi-slice excitation, the existing methods used a composite excitation pulse, yielded from the linear combination of radio frequency (RF) pulses with the same amplitude but different phases. The performance of such composite pulse, however, is limited, due to the square relationship between the peak pulse power and slice number. To solve this problem, we proposed a numerical optimization method for multi-slice excitation RF pulse based on both spin dynamics and optimal control theory. Together with slice selection gradient, the optimized pulse achieved selective excitation with arbitrary thickness, distance and number of slices. The performance of the optimized pulse in simultaneous multi-slice MRI was validated by simulation on multimodal imaging-based detailed anatomical (MIDA) digital phantom.

Key words: simultaneous multi-slice MRI, optimal control, numerical optimization, excitation profile