波谱学杂志

• •    

基于Simulink的连续波电子顺磁共振信号处理与检测仿真研究

陈博1,3,刘泉1,3,马蕾1,2,陈淑年1,贾亚琦1,4,朱斌3#,郭俊旺1,2*   

  1. 陈博1,3,刘泉1,3,马蕾1,2,陈淑年1,贾亚琦1,4,朱斌3#,郭俊旺1,2*
  • 收稿日期:2024-10-24 修回日期:2024-12-18 出版日期:2024-12-20 在线发表日期:2024-12-20
  • 通讯作者: 朱斌;郭俊旺 E-mail:guojunwang.thaa@vip.163.com;412833157@qq.com

Simulink-Based Simulation Study of Continuous Wave Electron Paramagnetic Resonance Signal Processing and Detection

CHEN Bo1,3,LIU Quan1,3,MA Lei1,2,CHEN Shunian1,JIA Yaqi1,4,ZHU Bin3#,GUO Junwang1,2*   

  1. CHEN Bo1,3,LIU Quan1,3,MA Lei1,2,CHEN Shunian1,JIA Yaqi1,4,ZHU Bin3#,GUO Junwang1,2*
  • Received:2024-10-24 Revised:2024-12-18 Published:2024-12-20 Online:2024-12-20
  • Contact: ZHU Bin;GUO Junwang E-mail:guojunwang.thaa@vip.163.com;412833157@qq.com

摘要: 根据连续波电子顺磁共振(CW-EPR)系统信号传输与检测的过程,通过Simulink平台构建了CW-EPR系统信号传输、调制与检测的仿真模型.该仿真模型涵盖了信号源模拟、调制、肖特基二极管检波器的检波以及基于锁相放大器的解调过程.基于该模型对调制幅度和调制相位变化下的DMPO样品波谱信号变化过程进行了评估.仿真结果表明针对不同的调制幅度和调制相位,波谱信号的仿真结果与实际测量结果保持一致.因此,本文建立的CW-EPR信号传输与检测仿真模型为深入理解CW-EPR现象、优化实验参数以及指导CW-EPR实验提供了理论支撑,可为后续研究中EPR系统设计与优化提供参考.

关键词: 连续波电子顺磁共振, 系统仿真, 波谱信号处理与检测, Simulink

Abstract: Based on the process of signal transmission and detection of the continuous wave electron paramagnetic resonance (CW-EPR) system, a simulation model of signal transmission, modulation, and detection of the CW-EPR system is constructed by the Simulink platform. The model covers the signal source simulation, modulation, detection by the Schottky diode detector, and demodulation process based on a lock-in amplifier. Based on the model, the spectral signal variation process of DMPO samples under modulation amplitude and modulation phase variations is evaluated. The simulation results show that for different modulation amplitudes and modulation phases, the simulation results of the wave spectrum signals are consistent with the actual measurements. Therefore, the simulation model of CW-EPR signal transmission and detection established in this paper provides theoretical support for the in-depth understanding of the CW-EPR phenomenon, the optimization of experimental parameters, and the guidance of CW-EPR experiments, and it can be used as a reference for the design and optimization of the EPR system in the subsequent research.

Key words: continuous wave electron paramagnetic resonance, system simulation, spectrum signal processing and detection, Simulink